Support til Astral

Produkt koder:
Astral 100 (27081)
Astral 150 (27083)

< Tilbage til produktside

Skift fra natmaske til mundstykke om dagen


Forstå Astral’s information bjælke

Aktiver Astral’s store knap funktion

Technical specifications

Continuous or intermittent ventilatory support for patients weighing more than 5 kg.*

  • Single circuit with leak
  • Single circuit with valve
  • Double circuit**
  • Mouthpiece circuit
  • Valve circuit therapy: CPAP, (A)CV, P(A)CV, P-SIMV, V-SIMV, PS
  • Leak circuit therapy: CPAP, (S)T, P(A)C, iVAPS* and optional AutoEPAP
  • Mouthpiece ventilation therapy: (A)CV, P(A)CV, PS
  • Manual breath**
  • Sigh breath (recruitment)**
  • Apnoea ventilation: Volume and pressure
  • 4 Preset programmes***
  • Delivered pressure (cm H2O):
    2–50 (leak circuit)
    0–50 (valve circuit)
  • Delivered tidal volume (mL):
    Adult: 100–2500 mL
    Paediatric: 50–300 mL
  • Respiratory rate:
    Adult: Off, 2–50 bpm
    Paediatric: Off, 5–80 bpm
  • SpO2 monitoring – Optional
  • FiO2 monitoring****
  • On screen:
    30 days of summary data
    Real-time pressure and flow (25 Hz)
  • ResScan (via download):
    365 days of summary data
    7 days of detailed data

A wide range of settable alarms is available including FiO2 and SpO2, when connected.

  • Weight: 3.2 kg
  • Physical and general specifications (L x W x H): 285 mm x 215 mm x 93 mm
  • Internal battery: Lithium-Ion battery, 14.4 V, 6.6 Ah, 95 Wh. 8 hours with a new battery under normal conditions – not intended to serve as a primary power source.*****
  • Sound power level: 43 ± 3 dBA †
  • Sound pressure level: 35 ± 3 dBA †
  • IEC 60601-1 classifications:
    Medical electrical equipment – general requirements for basic safety and essential performance
  • Class II (double insulation) Type BF
    Continuous operation
    Suitable for use with oxygen

Spørgsmål?